EasyManua.ls Logo

Adixen ASM 142 D - Set Point Menu

Adixen ASM 142 D
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
/5
C
140
Summary of
screens
value/function adjustable
Set
point menu
Level
1
SPECTROMETER
FILAMENT#1 :
on
TRIODE PRESSURE :
1
MARGIN :
100%
GB 01671 -
Edition
04 -
June
08
READY FOR
CYCLE
INLET VENT:
off
Level
2
SPECTROMETER
FILAMENT#1 :
on
TRIODE PRESSURE :
1
MARGIN :
100%
Level
3
E
400
Level
4
SPECTROMETER
FILAMENT#1 :
on
TRIODE PRESSURE :
1
MARGIN :
100%
E
400
NEXT
AUTOCALIBRATION off AUTOCALIBRATION
on
AUTOCALIBRATION
on
NEXT
OR
C
302
CAL. CHECK.:
operator
OR
CAL. CHECK.:
automatic
EVERY : 0100
CYCLES
EVERY : 0010
HOURS
FILAMENT CHOICE
:#1
FILAMENT :
on
CALIB. VALVE :
close
ELECTRONIC ZERO :
77
NEXT
TARGET
VALUE:1.0E-07
ACC. VOLTAGE :
220V
EMISSION :
0.6mA
COEF. SENS. :
01.00
OR
(1)
OR
FILAMENT CHOICE
:#1
FILAMENT :
on
CALIB. VALVE :
close
ELECTRONIC ZERO :
77
TARGET
VALUE:1.0E-07
ACC. VOLTAGE :
220V
EMISSION :
0.6mA
COEF. SENS. :
01.00
(2)
(1) Calibration = on
(2) Calibration = off
NEXT
(1) (2)
C
430
SELECTED GAS :
He
F2 => CALIBRATED
LEAK
INT.
CALIBRATED GAS :
He
F2
UNIT :
mbar.l/s
LEAK VALUE :
1.0E-07
LOCATION :
internal
NEXT
YEAR :
2003
LOSS PER YEAR : 2.00
%
REF TEMPERATURE : 20
¡C
TEMP. COEFF. : 3
%/¡C
NEXT
NEXT
C 305 C
305

Table of Contents

Other manuals for Adixen ASM 142 D