200
ASM 142 detector operating principle
GB 00208
-
Edition 03
- June
08
Vacuum
circuit
1
2
V
T3
3
VA1
7
18
V
C
V
T2
5
8
V
S
4
V
R1 VR2
VT1
Bacosol
valve
Minisol valves
15
Reference correspondence between
v
a
l
v
e
/
v
a
c
uum
block marks
E
530
1 Preamplifier
2 Analyzer cell
3
Detector
inlet port
4 Internal calibrated leak
5 Inlet
pressure
gauge
7 Vent connector
8
Long
distance sniffer connector
15
Roughing
primary pump
(RVP
20 0)
18
Detection molecular
pump (AMP
007I)
V
A1 Inlet vent valve (6 points)
V
R1
Roughing
valve ( point)
V
R2
Roughing
valve (2 points)
V
T1
Exhaust
valve (3 points)
V
T2
Detection
valve (4 points)
V
T3
Detection
valve (5 points)
V
S
Sniffing valve (9 points)
V
C
Calibration valve (7 points)