Adixen ASM 142 D - Sniffer Probe Clogged Reject Point; Starting a Sniffing Test; Ending a Sniffing Test

Adixen ASM 142 D
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
211
Operation of the leak detector
4
/
4
READY FOR CYCLE
INLET: ...
SNIFFING MODE
READY FOR CYCLE
INLET ...
GB 01362 -
Edition
02 -
September
04
Sniffing test
Starting a sniffing test
While the leak
detector
is in
stand-by
mode,
connect
the sniffer
probe
(accessory
to be purchased
separately A 700) to the sniffer port
of the leak detector.
STDBY
SNIFFING MODE
PLEASE WAIT ...
The
sniffing mode message
appears on the alphanumeric
display
.
The
sniffing
test
mode is operational.
Leak
value display
Ending
a sniffing test
Displays
are the
same
in hard
vacuum
and sniffing
test
modes.
Please
refer to hard
vacuum test
mode for the displays.
STDBY
Sniffer probe clogged
reject point
Accessories A
700
Advice
Block the sniffer probe end from time to time with a finger to check
that the helium signal goes down. If not, the probe may be clogged.
Sniffer probe
clogged
G
200