EasyManua.ls Logo

Adixen ASM 142 D - Calibration Checking Ticket with Internal Leak; Test Ticket

Adixen ASM 142 D
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B
320
Connecting the detector directly to
a
p
r
i
n
t
e
r
or
another
de
v
i
c
e
GB
00216
- Edition 02 - September
04
Calibration checking ticket
with internal
leak
5
C
A
L
I
B
R
A
T
I
O
N
I
N
F
O
R
M
A
T
I
O
N
S
:
DA
T
E
|
J
a
n
/
01
/
2003
T
I
M
E
|
01
:
19
:
23
current
i
n
t
e
r
n
a
l
t
e
m
pe
r
a
t
ur
e
(
C
)
|
24
current
c
o
e
f
.
s
e
n
s
|
00.78
g
l
o
b
a
l
r
a
t
e
|
1.10
E
-
07
background r
a
t
e
|
5.22
E
-
11
c
a
l
i
b
r
a
t
ed
l
e
a
k
-
r
a
t
e
|
1.10
E
-
07
t
a
r
ge
t
v
a
l
u
e
|
1.10
E
-
07
percent
a
ll
o
w
a
nc
e
(
+
/
-
)
|
15
R
E
S
U
L
T
(
%
)
|
-
0
T
est
ticket
DA
T
E
:
J
a
n
/
01
/
2003
H
O
U
R
CASE
PRESSURE
L
E
A
K
R
A
T
E
6
00:28:26
s
t
a
r
t
3.8
E
+
01
3.4
E
-
11
00:28:55
G
L
1.2
E
-
02
1.4
E
-
09
00:29:40
N
R
6.3
E
-
05
9.1
E
-
06
00:29:45
stop
5.8
E
-
05
N
R
1.1
E
-
05
Leak
value
DA
T
E
:
J
a
n
/
01
/
2003
H
O
U
R
CASE
PRESSURE
L
E
A
K
R
A
T
E
01:02:31
s
t
a
r
t
6.5
E
+
01
4.8
E
-
11
01:02:32
G
L
4.6
E
+
01
4.8
E
-
11
01:02:33
N
R
5.9
E
-
02
.8
E
+
00
01:02:36
stop
9.9
E
-
05
N
R
1.3
E
-
05
F
A
I
L
Leak
value
T
est
result if
Memo
function
active
Memo function C
550
5
/5

Table of Contents