44
Advanced
™
Osmometer Model 3250 User’s Guide
1. Set Block Bin #
This menu item displays the current block probe bin setting and allows
the setting to be changed.
The block probe bin setting should only need to be changed when
a new block probe of a different bin number is installed in your
instrument.
Upon entering Set Block Bin #, the current setting will be displayed. To
change the setting, enter your new block probe bin number by means
of the numeric keypad. Press ENTER to store the new setting or STOP
to restore the original setting. You may also use the > < keys to step
through the available choices. Once you press ENTER, you will be asked
to confirm the change. Press ENTER to confirm.
Changing the block probe bin number forces recalibration of the
instrument.
2. Set Sample Bin #
This menu item displays the current sample probe bin number setting
and allows the setting to be changed.
The sample probe bin number should only need to be changed when
a new sample probe of a different bin number is installed in your
instrument. Use the sample probe bin test procedure in the diagnostic
tests section (Chapter 5) of this user’s guide to determine the bin
number of a replacement probe.
Upon entering Set Sample Bin #, the current setting will be displayed.
To change the setting, enter your new sample probe bin number by
means of the numeric keypad. Press ENTER to store the new setting or
STOP to restore the original setting. You may also use the > < keys to
step through the available choices. Once you press ENTER, you will be
asked to confirm the change. Press ENTER to confirm.
Changing the sample probe bin number forces recalibration of the
instrument.