Osmometer Model 3250 User’s Guide
10. Select Language
This menu item enables selecting one of the alternate languages.
1. At Select Language, press START to display the current setting.
2. Press < or > to select one of the available languages.
3. Press ENTER to store the new setting (or STOP to restore the
original setting).
11. Product/Test
Product/Test is for factory use only and has no field function.
12. Data Capture
Pressing START at this SETUP menu item will bring you into an ON/
OFF selection menu.
When set to ON, the freezing curve output data will be presented to
the serial port independent of the system’s one-way hold feature. This
data is updated every 100ms after crossing 0°C and may be captured
with any RS-232 compatible computer program for use in plotting
freezing curves.
13. Assistance
Pressing START at this SETUP menu item will present you with
information on how to contact Advanced Instruments for supplies and
service. Important settings information which may be needed by our
service personnel will also be presented.
14. Set "Buzz" Amplitude
Note The amplitude of the freeze pulse has been optimized by the
factory. Adjustment of this setting is recommended only under
the supervision of an authorized service technician.
This menu item displays the current "buzz" amplitude setting and
allows the setting to be changed. The number displayed has no unit of
measure and may vary from assembly to assembly.