Troubleshooting and Service
There are noticeable signs of overheating or a burning odor.
The product does not operate normally when you follow the
operating procedures.
The main supply fuse(s) or any internal fuse(s) continually fail.
A discharge of static electricity from contact with the human body
or other conductor may damage system boards or static sensitive
devices. Never perform internal maintenance without following
recommended static protection procedures.
The product is equipped with operator accessible fuses. If a fuse
blows, it may be due to a power surge or failure of a component.
Replace the fuse only once. If the fuse blows a second time, it is
probably caused by failure of a component part. If this occurs, refer
service to qualified service personnel. Always replace the fuse with
one of the same rating, voltage, and type. Never replace the fuse
with one of a higher current rating.
When servicing the product, use only factory-specified parts.
The product contains an integral lithium battery that is not user-
serviceable.
WARNING When returning this product for service, or shipping this
product to a second location, remove all heat transfer fluid
and hazardous specimens and decontaminate the product
before packaging for shipment. If the product cannot be
decontaminated, consult with your shipping agent on
appropriate packaging and marking.