1200 Series Quarternary Pump User Manual 25
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tab le 3 Physical Specifications
Type Specification Comments
Weight 11 kg (25 lbs)
Dimensions
(width × depth × height)
140 x 345 x 435 mm (5.5 x 13.5 x 17
inches)
Line voltage 100 – 240 VAC, ± 10% Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5%
Power consumption 180 VA, 55 W / 188 BTU Maximum
Ambient operating
temperature
4–55 °C (41–131 °F)
Ambient non-operating
temperature
40–70 °C (-4–158 °F)
Humidity < 95%, at 25–40 °C (77–104 °F) Non-condensing
Operating Altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the module
Safety standards: IEC, CSA,
UL
Installation Category II, Pollution
Degree 2
For indoor use only. Research
Use Only. Not for use in
Diagnostic Procedures.