EasyManua.ls Logo

Agilent Technologies 1200 series - Exchanging the Purge Valve Frit or the Purge Valve

Agilent Technologies 1200 series
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1200 Series Quarternary Pump User Manual 85
Maintenance
7
Simple Repairs
Exchanging the Purge Valve Frit or the Purge Valve
1 Using a 1/4 inch wrench disconnect the pump outlet capillary at the purge
valve.
2 Disconnect the waste tube. Beware of leaking solvents due to hydrostatic
pressure.
3 Using the 14 mm wrench unscrew the purge valve and remove it.
4 Remove the plastic cap with the gold seal from the purge valve.
When Frit when plunger seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 ml/min of H
2
0 with purge valve opened)
Purge valve if internally leaking
Tools required Wrench 1/4 inch
Wrench 14 mm
Pair of tweezers or toothpick
Parts required # Part number Description
5 01018-22707 PTFE frit (pack of 5)
1 G1311-60009 Purge valve
Preparations Switch off pump at the main power switch
Remove the front cover

Table of Contents

Other manuals for Agilent Technologies 1200 series

Related product manuals