58 1200 Series Quarternary Pump User Manual
5 Optimizing Performance
When to use the Seal Wash Option
When to use the Seal Wash Option
Highly-concentrated buffer solutions will reduce the lifetime of the seals and
plungers in your pump. The seal wash option allows to maintain the seal
lifetime by flushing the back side of the seal with a wash solvent.
The seal wash option is strongly recommended when buffer concentrations of
0.1 Molar or higher will be used for long time periods in the pump.
The seal wash option can be ordered by quoting part number 01018-68722 (kit
contains all parts needed for one pump head). The active seal wash option kit
can be ordered by quoting part number G1311-68711.
The seal wash option comprises a support ring, secondary seal, gasket and
seal keeper for both plunger sides. A wash bottle filled with water
/isopropanol (90/10) should be placed above the pump in the solvent cabinet
and gravity will maintain a flow through the pump head removing all possible
buffer crystals from the back of the pump seal. For the active seal wash a
peristaltic pump is pumping the solvent through the pump head.
For information on the installation of the active seal wash option refer to
“Installing the Seal Wash Option on page 94.
NOTE
Running dry is the worst case for a seal and drastically reduces its lifetime. The seal will
build up sticky layers on the surface of the plunger. These sticky layers will also reduce the
lifetime of the primary seal. Therefore the tubes of the wash option should always be filled
with solvent to prolong the lifetime of the wash seal. Always use a mixture of bidistilled
water (90 %) and isopropanol (10 %) as wash solvent. This mixture prevents bacteria
growth in the wash bottle and reduces the surface tension of the water.

Table of Contents

Other manuals for Agilent Technologies 1200 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 series and is the answer not in the manual?

Agilent Technologies 1200 series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals