1200 Series Quarternary Pump User Manual 81
Maintenance
7
Simple Repairs
Exchanging the Active Inlet Valve Cartridge
Figure 14 Exchanging the Active Inlet Valve
1 Remove the front cover.
2 Unplug the active inlet valve cable from the connector.
When If internally leaking (backflow)
Tools required Wrench 14 mm
Pair of Tweezers
Parts required # Part number Description
1 G1312-60025 Active inlet valve body
1 5062-8562 Valve cartridge (400 bar)
Preparations Switch off pump at the main power switch and unplug the power cable
8dccZXidg
HdakZci^caZiijWZ
6YVeiZg
6Xi^kZ^caZikVakZ
8Ve^aaVgn
ejbeid^c_ZXi^dcYZk^XZ
6Xi^kZ^caZikVakZXVWaZ