90 1200 Series Quarternary Pump User Manual
7 Maintenance
Simple Repairs
3
Using one of the plungers carefully remove the seal from
the pump head (be careful not to break the plunger).
Remove wear retainers, if still present.
4 Insert new seals into the pump head.
5 Reassemble the Pumphead assembly.