EasyManua.ls Logo

Agilent Technologies 1260 Infinity II - Page 145

Agilent Technologies 1260 Infinity II
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity II Preparative Autosampler User Manual 145
Error Information
8
Sample Cooler Error Messages
Cooler temperature below or above target temperatur.
Error ID: 30705, 30706
The sample cooler temperature is above or below the target temperature.
Cooler Overpressure Failure
Error ID: 30712
The sample cooler pressure sensor has detected an overpressure.
Probable cause Suggested actions
1
Condensate inside of the chiller
Check if the drainage is ok.
2
Defective cable
Check the signal cable from the ALS to the
sample cooler.
3
Defective temperature sensor
Replace the sample cooler.
Probable cause Suggested actions
1
Overpressure in the chiller
Switch off the instrument and wait for 15 min.
Restart the instrument and check the signals in
Lab Advisor.
2
Defective compressor
Replace the sample cooler.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity II

Related product manuals