146 Agilent 1260 Infinity II Preparative Autosampler User Manual
8
Error Information
Sample Cooler Error Messages
Cooler Condensate Error
Error ID: 30702, 30709, 30710
The sample cooler has detected condensate inside of the leak pane.
Probable cause Suggested actions
1
Blocked drainage
Switch off the instrument and remove the 
condensate tubings. Place a beaker underneath 
the waste outlet. Check the waste container. 
There should be no kinked and sealed tubing. 
For further information see “Install the Sample 
Cooler” on page 60.
2
Defective condensate sensor
• Switch the sample cooler off/on and check 
if after 8 min the error shows up again.
• Replace the condensate sensor.
3
Defective leak sensor
Replace the sample cooler.