92 Agilent 1260 Infinity II Preparative Autosampler User Manual
5
Preparing the Module
Setting up the Sampler
Table 9 Method parameter settings
Injection
Specify the injection volume (in µL) in the Injection 
Volume field. The injection volume limits depend upon the 
configuration settings of the autosampler. Note that this 
setting is ignored when Pretreatment/Injector Program is 
defined.
Needle Wash
Mark the Enable Needle Wash check box to specify an 
injection with needle wash.
The following wash parameters can be specified:
• Mode: Select the wash location by clicking the down 
arrow and selecting the needle wash mode from the 
list. You can select from Flush Port (default location) or 
Wash Vial.
•Time: This field is enabled only if you have selected 
Flush Port. Specify a flush time in seconds between 
1 and 100 in steps of 1.
•Location: enabled only if you have selected a Wash 
Vial. Allows you to specify a wash location in the 
format Vial X, where X is the vial number. Specify well 
locations in the format P<Plate #>-<Row 
Character><Column Number>, for example, P1-A1. The 
values of <Row Character> and <Column Number> 
depend on the wellplate type.
• Repeat: available only if you have selected Wash Vial. 
You can specify up to 10 repeat washes.
Stoptime: enables you to set the time that the analysis 
stops. Limits: 0.01 through 99999 min or As Pump/No 
Limit.
Posttime: the Vialsampler remains in a post-run state 
during the Posttime to delay the start of the next analysis. 
A Posttime period can be used to allow the column to 
equilibrate after changes in solvent composition (for 
example after gradient elution). Limits: 0.01 through 
99999 min or Off.