EasyManua.ls Logo

Agilent Technologies 16902A

Agilent Technologies 16902A
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
109
Chapter 5: Troubleshooting
d Select Close to close the pod assignment window.
e Under the Format tab, select the field showing the channel assignment
for the pod under test, (you may need to use the scroll bar) then select
“****************” to activate all channels.
f Select the Apply button at the bottom of the window.
7 Verify the logic analyzer threshold setting
a Under the Format tab, select the CLK Thresh... field (you may need to
use the scroll bar). The Clock Thresholds window will appear.
b Ensure that the threshold field associated with J-clock says Differential.
c If it doesn’t, ensure the correct probe (E5379A) is attached to the pod.
The threshold is set to Differential automatically when the E5379A
probe is attached.
d Select Close to close the Clock thresholds window.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals