TIP
The fragment table can be directly edited in the setpoint explorer:
NOTE
Changing the start or end points of the fragment will change the calculated rRNA ratio.
It might be convenient to pause the automatic analysis (Electropherogram > Pause
Automatic Analysis) until all changes are done.
Setting the Baseline for Calculation of RNA Concentration
At low signal-to-noise ratios, the baseline that defines the area used for calculating the
concentration of RNA assays is highly dependent on the settings for the Start and End
Time. You can adjust the Start and End Times manually (thereby adjusting the baseline)