• Reassign lower marker
DNA and Protein Assays Only:
• Exclude peaks from analysis
• Reassign upper/lower markers
• Alignment or no alignment with marker peaks
• Manual integration
Protein Assays Only:
• Absolute quantitation
TIP
When applying modified data analysis setpoints, the software will (by default)
immediately recalculate the raw data, which takes some time. Select Don’t Analyze
from the Gel Menu or Electropherogram Menu to temporarily switch off automatic
data analysis while you modify setpoints.
If you save the data file after making changes, it will keep a record of the changes such
as gel color, sample names, and peak find settings that were in effect at the time the file
is resaved. If you do not want to change the original file, choose File > Save As… and give
the file a new name or save it to a different location.