b Select the Data Acquisition Mode.
Select Default, if you want the measurement time to be set to the default value
(240 s/sample). The maximum time is shown in brackets.
– OR –
Select Fixed time and enter the time in [s] that the measurement of each sample is
to take.
– OR –
Select Number of Cells and enter the minimum number of events that should be
measured. In the and no longer than field, enter the maximum time in [s] a
measurement can take, regardless of whether or not the defined number of events is
reached. The maximum time is shown in brackets.
NOTE
The overall run time for a chip is limited to 1440 s. The individual run time for one
sample depends on the number of samples that are measured per chip. If only one
sample is measured, you can set the run time up to 1440 s.