6 Select a Destination for the chip data file (.xad) that will be generated as the result of
the chip run. You can also specify a custom File Prefix for this file.
7 Under Data Acquisition Parameters, enter the number of samples you want to have
measured.
The total number of samples that can be measured varies with the type of assay
selected. With DNA and RNA Nano assays, 12 samples may be run; with RNA Pico
assays, 11 samples may be run; and with Protein assays, the maximum number of
samples is 10. When preparing the chip (see “Preparing Samples, Reagents, and Chips
for Electrophoretic Assays” on page 54), keep in mind that you have to follow the
sequence of the sample wells. For example, if you want to measure only 3 samples, you
have to fill the wells 1, 2, and 3 of your chip.