*RCL {0|1|2|3}
Recall a previously stored state. To recall a stored state, you must use
the same memory location used previously to store the state.
You cannot recall the instrument state from a memory location that
was not previously specified as a storage location. For example, an
error is generated if you attempt to recall from memory location2
but have never stored to that location. From the remote interface, a
+810, “State has not been stored” error is generated if nothing is stored
in the specified memory location.
MEMory:STATe:DELete {0|1|2|3}
Delete a previously stored state and clear the memory location.
If nothing is stored in the specified memory location, a +810, “State has
not been stored” error is generated.
4
Chapter 4 Remote Interface Reference
Output Configuration Commands
153