RS-232 Interface Commands
Use the front-panel I/O MENU to select the baud rate, parity, and
number of data bits (see pages 219 and 220 for more information).
SYSTem:LOCal
Place the function generator in the local mode for
RS-232 operation.
All keys on the front panel are fully functional.
SYSTem:REMote
Place the function generator in the remote mode for
RS-232 operation.
All keys on the front panel, except the
LOCAL key, are disabled.
It is very important that you send the SYSTEM:REMOTE command
to place the function generator in the remote mode. Sending or receiving
data over the RS-232 interface when not configured for remote operation
can cause unpredictable results.
SYSTem:RWLock
Place the function generator in the remote mode for
RS-232 operation.
This command is the same as the SYSTEM:REMOTE command except
that all keys on the front panel are disabled, including the
LOCAL key.
Ctrl-C
Clear the operation in progress over the
RS-232 interface and discard
any pending output data. This is equivalent to the IEEE-488 device clear
action over the GPIB interface.
Chapter 4 Remote Interface Reference
RS-232 Interface Commands
200

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals