EasyManua.ls Logo

Agilent Technologies 33120A - RS-232 Interface Commands

Agilent Technologies 33120A
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RS-232 Interface Commands
Use the front-panel I/O MENU to select the baud rate, parity, and
number of data bits (see pages 219 and 220 for more information).
SYSTem:LOCal
Place the function generator in the local mode for
RS-232 operation.
All keys on the front panel are fully functional.
SYSTem:REMote
Place the function generator in the remote mode for
RS-232 operation.
All keys on the front panel, except the
LOCAL key, are disabled.
It is very important that you send the SYSTEM:REMOTE command
to place the function generator in the remote mode. Sending or receiving
data over the RS-232 interface when not configured for remote operation
can cause unpredictable results.
SYSTem:RWLock
Place the function generator in the remote mode for
RS-232 operation.
This command is the same as the SYSTEM:REMOTE command except
that all keys on the front panel are disabled, including the
LOCAL key.
Ctrl-C
Clear the operation in progress over the
RS-232 interface and discard
any pending output data. This is equivalent to the IEEE-488 device clear
action over the GPIB interface.
Chapter 4 Remote Interface Reference
RS-232 Interface Commands
200

Table of Contents

Other manuals for Agilent Technologies 33120A

Related product manuals