Self-Test Errors
The following errors indicate failures that may occur during a self-test.
Refer to the Service Guide for more information.
590 I/O processor reset; possible low power line voltage
601 Front panel does not respond
602
RAM read/write failed
603 Waveform
RAM readback failed
604 Modulation
RAM readback failed
605 Serial configuration readback failed
606 Waveform
ASIC failed
607
SYNC signal detection failure, Bessel filter path
608
SYNC signal detection failure, Elliptic filter path
625 I/O processor does not respond
626 I/O processor failed self-test