The Standard Event Register
The Standard Event register reports the following types of instrument
events: power-on detected, command syntax errors, command execution
errors, self-test or calibration errors, query errors, or when an *OPC
command is executed. Any or all of these conditions can be reported in
the Standard Event summary bit through the enable register. To set the
enable register mask, you write a decimal value to the register using the
*ESE (event status enable) command.
An error condition (Standard Event register bits 2, 3, 4, or 5) will always
record one or more errors in the function generator’s error queue.
Read the error queue using the SYSTem:ERRor? command.
Bit Definitions – Standard Event Register
Bit
0 Operation Complete
1 Not Used
2 Query Error
3 Device Error
4 Execution Error
5 Command Error
6 Not Used
7 Power On
Decimal
Value
1
2
4
8
16
32
64
128
Definition
All commands prior to and including an *OPC
command have been executed.
Always set to 0.
The function generator tried to read the output buffer
but it was empty. Or, a new command line was
received before a previous query has been read.
Or, both the input and output buffers are full.
A self-test or calibration error occurred (see error
numbers 501 through 857 in chapter 5).
An execution error occurred (see error numbers
-211 through -224 in chapter 5).
A command syntax error occurred (see errors
-101 through -168 in chapter 5).
Always set to 0.
Power has been turned off and on since the last
time the event register was read or cleared.
4
Chapter 4 Remote Interface Reference
The SCPI Status Registers
207