-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for
the command. You may have used an invalid parameter choice.
Example: DISP:STAT XYZ (XYZ is not a valid choice).
-330 Self-test failed
The function generator’s self-test failed from the remote interface
(*TST? command). In addition to this error, one or more specific errors
may be reported. See also “Self-Test Errors,” on page 237.
-350 Too many errors
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared when power has been off, or after a *CLS
(clear status) command has been executed.
-410 Query
INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous
data is not overwritten). The output buffer is cleared when power has
been off, or after a device clear has been executed.
-420 Query
UNTERMINATED
The function generator was addressed to talk (i.e., to send data over the
interface) but a command has not been received which sends data to the
output buffer. For example, you may have executed an APPLy command
(which does not generate data) and then attempted an ENTER statement
to read data from the remote interface.
-430 Query
DEADLOCKED
A command was received which generates too much data to fit in the
output buffer and the input buffer is also full. Command execution
continues but all data is lost.
-440 Query
UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a
command string. Example: *IDN?;:SYST:VERS?
5
Chapter 5 Error Messages
Execution Errors
235