You can use the *OPC? (operation complete query) command or the
*OPC (operation complete) command to signal when the sweep is
complete. The *OPC? command returns “1 to the output buffer when
the sweep is complete. The *OPC command sets the “operation complete”
bit (bit 0) in the Standard Event register when the burst is complete.
TRIGger:SOURce?
Query the present trigger source. Returns
IMM”,EXT”, orBUS”.
4
Chapter 4 Remote Interface Reference
Frequency Sweep Commands
173