EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>4100

Agilent Technologies 4100 Site Preparation Guide

Agilent Technologies 4100
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Laboratory Environment
20 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Environmental Conditions
The area selected for the operation of an Agilent 4100 MP-AES
instrument must be free from drafts, corrosive atmospheres and
vibration. Sample preparation areas and materials storage facilities
should be located in a separate room.
The area should be a dust free, low humidity environment. Air-
conditioning is strongly recommended for control of the
environment.
Table 1. Suitable environmental conditions
Condition Altitude Temp (°C) Humidity
(%RH) non-
condensing
Max Rate Of
Change (°C)
Non-operating
(Transport)
0–4,000 m
(0–13100 ft)
-40–70 °C
(-40–158 °F)
<90
Operating and
complying with
performance
specifications
0–3,000 m
(0–9,800 ft)
5–35 °C
(41–95 °F)
50–80 2 °C /hr
±10°C since
calibration
3,000–4,000 m
(9,800–13,100 ft)
5–25 °C
(41–77 °F)
50–80 2 °C /hr
±10 °C since
calibration
NOTE
For optimum analytical performance, it is recommended that the ambient
temperature of the laboratory be between 20–25 °C (68–77 °F). The ambient
temperature must be constant, with the maximum acceptable temperature
change being 2 °C (3.6 °F) per hour to comply with performance specifications.
The maximum temperature change before a recalibration is required is
±10 °C
(±18 °F).

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals