Accessories and Options
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 37
6. Accessories and Options
Organic Solvents or Hydrofluoric Acid 37
External Gas Control Module 38
Monochromator Optics Air Purge 39
Sample Preparation System (SPS 3) 40
Multimode Sample Introduction System (MSIS) 41
Agilent 4107 Nitrogen Generator 42
This section includes site preparation information on Agilent
manufactured accessories suitable for use with Agilent 4100 MP-AES.
The following information should be used to supplement the
documentation included with each accessory.
Organic Solvents or Hydrofluoric Acid
If organic solvents or hydrofluoric acid are intended to be used, then
special organic resistant pump tubing is highly recommended.
For analysis of hydrogen fluoride (HF) containing solutions, a
purpose designed Ezyfit plasma torch with ceramic injector and
specialized sample introduction system consisting of polymer
nebulizer and spray chamber are required.
Please refer to the Agilent website for details and part number
information: http://www.chem.agilent.com/