Laboratory Facilities
32 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Other electrical connections
USB
NOTE
Basic insulation is provided for single fault protection on the USB connector.
Fuses
1CB1 Circuit breaker 12 A fast trip
NOTE
For safety reasons, any other internal fuse or circuit breaker is not operator
accessible, and should only be replaced by Agilent-authorized personnel.
Gas Supplies
The installation of compressed or liquid gas supplies must comply
with the rules and/or regulations imposed by the local authorities
responsible for such use in the workplace.
Depending on the instrument configuration and accessories, up to
five different gas supplies are required such as nitrogen and argon
for plasma generation, nitrogen for purging and air for the nitrogen
generator and for some accessories.