Accessories and Options
40 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Gas requirements
All gas supply installations must comply with rules and/or
regulations that are imposed by the local authorities responsible for
the use of compressed gas in the workplace.
Table 8 Monochromator Optics Air Purge
Air
Quality
ISO 8573-1:2010 CLASS 1.4.3
Permissible pressure range 450–600 kPa (58–87 psi)
Recommended pressure 600 kPa (87 psi)
Required flow rates 10 SLPM
NOTE
Failure to use the specified air quality will significantly may result in
malfunction or damage to the 4100 MP-AES.
Sample Preparation System (SPS 3)
The Agilent Sample Preparation System 3 (SPS 3) is compatible with
a wide range of commercially available low-cost autoclavable sample
racks.
Sample contamination from airborne particles is eliminated and
corrosive or toxic fumes are removed during sampling with the
optional environmental enclosure.

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals