Accessories and Options
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 39
Table 7. EGCM Gas specifications
Nitrogen Air
Purity
99.95 % n/a
Oxygen
< 0.05 % n/a
Water vapor
< 4 ppm n/a
Quality
n/a ISO 8573-1:2010 CLASS
1.4.3
Permissible pressure range
450–600 kPa (65–87 psi) 450–600 kPa (65–87 psi)
Recommended pressure
500 kPa (72.5 psi) 600 kPa (87 psi)
Required flow rates
10 SLPM 1.5 SLPM
NOTE
A nitrogen generator may be used to supply nitrogen to the ECGM. However, it
must be a separate unit from that used for the instrument.
Monochromator Optics Air Purge
(Part number G8000-68001)
The monochromator is a precision mechanical device and its
continued stable and reproducible operation relies on clean
corrosion free surfaces.
The Monochromator Optics Air Purge accessory is designed to
minimize the ingress of particles and corrosive vapors, commonly
encountered in the typical laboratory, into the monochromator by
providing a continuous flow of clean, dry, oil-free and dust free air
into the monochromator assembly. The use of this accessory provides
additional protection for your instrument.

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals