Gas Requirements and Configurations
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 45
7. Gas Requirements and
Configurations
Compressed Air Requirements 45
Agilent 4100 MP-AES Gas Configurations 48
Compressed Air Requirements
It is recommended to consult a professional compressed air supplier
to ensure that your compressed air quality meets the ISO
requirements. Most basic industrial or domestic installations will not
meet these requirements. Using compressed air that does not meet
Agilent’s specification may result in damage or malfunction of the
Agilent 4100 MP-AES or Agilent 4107 Nitrogen Generator.
Two different qualities of compressed air are required for the Agilent
4100 MP-AES.
The air quality for the Agilent 4107 Nitrogen Generator will typically
require an oil free compressor followed by a dryer stage to remove
water from the compressed air. The Nitrogen Generator is supplied
with a regulator/filter assembly to ensure particulate matter is
removed.
The compressed air for the pre-optics protection (POP) Gas, EGCM
and Monochromator Air Purge will require additional filtration after
the drying stage. This is the responsibility of the customer as part of
the compressed air system installation.

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals