Gas Requirements and Configurations
46 Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Table 13. Air compressor requirements
Equipment Quality
Specification
Pressure
Nitrogen Generator ISO 8573-1:2010
CLASS 8.4.3
620–750 kPa (90–109 psi)
POP, EGCM,
Monochromator Air
Purge
ISO 8573-1:2010
CLASS 1.4.3
450–600 kPa (58–87 psi)
CAUTION
The nitrogen generator will be permanently damaged by oil vapor or aerosol from
a compressor. The damage will not be repairable and will require replacement of
the product
CAUTION
Oil-free compressors may still deliver oil (hydrocarbons) to the nitrogen generator
if the inlet air is contaminated. It is essential that clean air, free of hydrocarbons
and acid, is supplied to the inlet of the compressor. A three-stage filter is used
between the compressor and nitrogen generator to protect the nitrogen
generator from any remaining contamination.
CAUTION
Even when an oil-free compressor is used the filter unit incorporated in the
nitrogen generator must be used.