Laboratory Facilities
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide 29
5. Laboratory Facilities
Exhaust System 29
Electrical Power Supplies 30
Electrical requirements 31
Instrument power connections 31
Other electrical connections 32
Fuses 32
Gas Supplies 32
Cylinder storage instructions 34
Cryogenic liquids 34
Waste Fluid Container 35
Guidelines for Software Installation 35
This section includes information on the laboratory facilities
required to support the operation of the instrument.
Exhaust System
The plasma operates at a temperature of approximately 6,000 °C,
(11,000 °F) with up to 1,000 Watts of power (6,480 kilojoules per
hour). The instrument exhaust fumes can be noxious or corrosive.
The Agilent 4100 MP-AES instrument chimney must be connected to
an exhaust system. The exhaust system must include an exhaust fan,
be ducted to an external vent and provide a minimum flow of
3 m
3
/min (106 cfm) at the intake point.
The exhaust system installation must comply with any rules and/or
regulations that may be imposed by the local authorities responsible
for control of facilities and fixtures in the work place.

Other manuals for Agilent Technologies 4100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4100 and is the answer not in the manual?

Agilent Technologies 4100 Specifications

General IconGeneral
BrandAgilent Technologies
Model4100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals