EasyManua.ls Logo

Agilent Technologies 4349B - Check Procedure References; Check 1, DC-DC Converter Output Voltage; DC-DC Converter Output Voltage Check; DC-DC Converter Output Voltages (Connected)

Agilent Technologies 4349B
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Check
Procedure
References
Refer
to
the
corresponding
check
procedure
.
Each
check
procedure
is numbered
the same
as in
Figure
4-7
for
quick
reference
.
Check
1,
DC-DC
Converter
Output
V
oltage
1.
Remove
the
cover
referring
to
\Cover
Removal"
in
Chapter
3.
2.
Turn
on
the
4349B
.
3.
Using
a
multimeter
,
check
to
see
if the
voltages between
the test
pins
(TP
s),
shown
in
T
able
4-1
,
on
A1
Main
Board
Assembly are
in the
ranges shown
in
T
able
4-1
.
4.
If
passed,
proceed
to
\Check
2,
LCD
Displays Anything?",
otherwise proceed
to \Check
9,
Fuse
and
Line
Switch
".
Figure 4-8. DC-DC Converter Output V
oltage Check
Table 4-1. DC-DC Converter Output V
oltages (Connected)
Test Pin (+) T
est Pin (
0
) Range
11 15 5.1
6
0.2 V
13 15 12
6
0.6 V
14 15
0
12
6
0.6 V
Troubleshooting 4-11

Table of Contents

Other manuals for Agilent Technologies 4349B

Related product manuals