EasyManuals Logo

Agilent Technologies 5100 ICP-OES Site Preparation Guide

Agilent Technologies 5100 ICP-OES
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
Safety Practices and Hazards
Agilent 5100 and 5110 ICP-OES Site Preparation Guide 11
Heat, Vapors and Fumes
Heat, ozone, vapors and fumes generated by the plasma can be
hazardous, and must be extracted from the instrument by means of
an exhaust system. Ensure that an exhaust system of the appropriate
type is fitted (see Page 36). The system must be vented to the outside
air in accordance with local regulations and never within the
building. Regularly check the exhaust system by smoke test to ensure
that the exhaust system is functioning correctly. The exhaust fan
must always be switched on before igniting the plasma.
Compressed Gas Hazards
All compressed gases (other than air) can create a hazard if they leak
into the atmosphere. Even small leaks in gas supply systems can be
dangerous. Any leak (except that of air or oxygen) can result in an
oxygen-deficient atmosphere, which can cause asphyxiation. The
area in which cylinders are stored and the area surrounding the
instrument must be adequately ventilated to prevent such gas
accumulations.
Gas cylinders must be stored and handled strictly in accordance with
local safety codes and regulations. Cylinders must be used and stored
only in a vertical position and secured to an immovable structure or
a properly constructed cylinder stand. Move cylinders only by
securing them to a properly constructed trolley.
Use only approved regulator and hose connectors (refer to the gas
supplier’s instructions). Keep gas cylinders cool and properly
labeled. (All cylinders are fitted with a pressure relief device that will
rupture and empty the cylinder if the internal pressure is raised
above the safe limit by excessive temperatures.) Ensure that you have
the correct cylinder before connecting it to the instrument.
The primary gas to be used with the spectrometer is argon, which is
the conductive gas for the plasma. Argon or nitrogen can be used as
the polychromator purge gas. Other gases may be required for future
options and accessories. Use only ‘instrument grade’ gases with your
spectrometer.
If using cryogenic gases (for example, liquid argon), prevent severe
burns by wearing suitable protective clothing and gloves.

Table of Contents

Other manuals for Agilent Technologies 5100 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 ICP-OES and is the answer not in the manual?

Agilent Technologies 5100 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5100 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals