Laboratory Facilities
Agilent 5100 and 5110 ICP-OES Site Preparation Guide 41
If necessary, replace the power cord only with a cord equivalent to
the one specified (See Appendix A).
CAUTION
If an instrument is being ordered from one location, but is to be installed in
another location with different electrical power characteristics, this must be
noted on the order. A special note must also be made if the electrical power at
the site is different from the standard electrical power in that country.
WARNING
Make sure the power cords supplied with the Agilent ICP-OES are
appropriate for your country and site before using them.
Other Connections
IEEE 802.3, Ethernet LAN cable
After the electrical requirements have been met, check the checklist box: Specified
electrical supply and power outlets installed.
Circuit Breaker
NOTE
For safety reasons, any internal fuse or circuit breaker is not operator accessible,
and should only be replaced by Agilent-authorized personnel.
The mains power switch contains a 20 A circuit breaker, which is
reset when the power switch is cycled.
Gas Supplies
The installation of compressed or liquid gas supplies must comply
with the rules and/or regulations imposed by the local authorities
responsible for such use in the workplace.