Laboratory Facilities
Agilent 5100 and 5110 ICP-OES Site Preparation Guide 43
Table 8. Typical flow rates for the Agilent ICP-OES instruments
Argon (with argon purge gas) Nitrogen (as purge gas)
Standby mode 0.70 L/min Nitrogen flow 0.8 L/min
Operational range min-max
(plasma on)
9.1-31.8 L/min Argon flow 8.4-28.1 L/min
Nitrogen flow 0.8-4.4 L/min
Typical flows
Measuring wavelengths
> 189 nm (poly boost off)
14.65-20.65 L/min Argon flow 13.95-19.95 L/min
Nitrogen flow 0.8 L/min
Measuring wavelengths
< 189 nm (poly boost on)
19.25-25.25 L/min Argon flow 15.55-21.55 L/min
Nitrogen flow 4.4 L/min
The Agilent ICP-OES is supplied with three PVDF gas supply hose
assemblies, each 3 m (9.8 ft) in length fitted with 1/4 inch Swagelok nut
and ferrule set.
Fittings for the regulator end must be supplied by the customer.
Alternatively, a fittings kit can be ordered from Agilent Technologies.
See the Agilent website at www.agilent.com for ordering information.
Customers will provide connection from the gas supply to a shutoff
valve.
Customers must provide pressure regulators for the gas cylinders.
When ordering, consider the size of the output tubing, 1/4 inch
(6.4mm), and also the Compressed Gas Association (CGA) number.
If the Nitrogen Purge option is purchased, additional fittings will be
supplied to connect the gas to the ICP-OES.
Storage Cylinder Instructions
Cylinders containing gas under pressure should be firmly secured to
a rigid structure, and the storage area must be adequately ventilated.
Never locate gas cylinders near a source of ignition, or in a position
that is subject to direct heat. Gas storage cylinders often incorporate
a pressure release device, which will discharge the gas at a
predetermined temperature, usually around 52 °C (125 °F).

Table of Contents

Other manuals for Agilent Technologies 5100 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 ICP-OES and is the answer not in the manual?

Agilent Technologies 5100 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5100 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals