Installation
26 Agilent 5800 and 5900 ICP-OES User's Guide
ICP-OES torches
The range of ICP-OES torches are listed below, each with unique properties to suit
a wide range of applications. Additional torches may be available. See the Agilent
website for more information.
Parameter
Dual View Torches
Standard torch
Semi-demountable
torch
Inert semi-
demountable torch
High solids semi-
demountable torch
Construction One piece
Two piece
demountable
Two piece
demountable
Two piece
demountable
Injector inner diameter
(mm) 1.8 1.8 1.8 2.4
Injector material Quartz Quartz Ceramic Quartz
Outer tube length Medium with slot Medium with slot Medium with slot Medium with slot
Outer tube material Quartz Quartz Quartz Quartz
Intermediate tube Quartz tulip Quartz tulip Quartz tulip Quartz tulip
Recommended
application
Most sample types Most sample types Hydrofluoric (HF)
acid digests
Samples with high
total dissolved solids
Parameter
Radial View Torches
Standard torch Semi-demountable torch Inert semi-demountable torch
Construction One piece Two piece demountable Two piece demountable
Injector inner
diameter (mm) 1.4 1.4 1.8
Injector material Quartz Quartz Ceramic
Outer tube length Short (no slot) Short (no slot) Short (no slot)
Outer tube material Quartz Quartz Quartz
Intermediate tube Quartz tulip Quartz tulip Quartz tulip
Recommended
application
Most sample types Most sample types Hydrofluoric (HF) acid digests