Maintenance and Troubleshooting
64 Agilent 5800 and 5900 ICP-OES User's Guide
When replacing the communications cable, always use a “Category 5E Patch Cord,
STP/FTP, Shielded, Stranded, White, 14 ft, Booted” rated Ethernet cable to maintain
EMC compliance.
The following is a list of recommended spares to keep on hand to minimize
downtime during maintenance and repairs:
Torch (one piece) or spare components such as the outer tube-set and
injector, if using a demountable torch
Air inlet filter element (dust filter)
Axial pre-optic window
Radial pre-optic window
Spray chamber
Nebulizer
Peristaltic pump tubing
Drain tubing
Technical Support
For technical support contact information, refer to the Agilent Technologies
website for details:
www.agilent.com