EasyManua.ls Logo

Agilent Technologies 5800

Agilent Technologies 5800
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Troubleshooting
64 Agilent 5800 and 5900 ICP-OES User's Guide
When replacing the communications cable, always use a “Category 5E Patch Cord,
STP/FTP, Shielded, Stranded, White, 14 ft, Booted” rated Ethernet cable to maintain
EMC compliance.
The following is a list of recommended spares to keep on hand to minimize
downtime during maintenance and repairs:
Torch (one piece) or spare components such as the outer tube-set and
injector, if using a demountable torch
Air inlet filter element (dust filter)
Axial pre-optic window
Radial pre-optic window
Spray chamber
Nebulizer
Peristaltic pump tubing
Drain tubing
Technical Support
For technical support contact information, refer to the Agilent Technologies
website for details:
www.agilent.com