Operation
46 Agilent 5800 and 5900 ICP-OES User's Guide
4 Click the Print button to generate a report as specified. A message will appear
briefly indicating the progress status of the report being generated.
5 The Print dialog box will then be opened. You can select to print all of the
report, or a range of pages. Reports are printed to your default printer, unless
you specify otherwise. You can specify your printer options in the Print Setup
dialog box, accessible from the File menu.
Turning Off the Instrument
There are three standby and shutdown modes:
Standby Mode with argon on – for in between routine use (Page 47)
Standby Mode with argon off – for overnight, weekend, and long-term
shutdown (Page 47)
Full shutdown – for a complete shutdown (Page 48)
Instrument Standby Mode with argon on
This Standby Mode option leaves the 5800/5900 ICP-OES in a state of readiness
while only consuming low amount of utilities (argon gas and electricity). The
benefit of being in Standby Mode is the 5800/5900 can be operational again in <20
minutes from plasma ignition. Standby Mode is automatically enabled once the
previous analysis is finished and the plasma is shut down. Follow ‘Setting the
instrument to Standby Mode with argon on or off’ on Page 47 to ensure the
instrument is fully prepared for Standby Mode.
Typical recommendation for Standby Mode is to leave the ICP-OES power on and
the argon gas supply pressured up, allowing a trickle purge of argon through the
polychromator, while not collecting analytical data.
To get back to operational status from this mode, turn on the plasma and stabilize
for 20 minutes.
Instrument Standby Mode with argon off
For overnight, weekend, and long-term shutdown is recommended the instrument
is fully powered but the plasma off. In this mode the purge systems are also
turned off. The internal air cooling system runs at reduced capacity.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5800 and is the answer not in the manual?

Agilent Technologies 5800 Specifications

General IconGeneral
BrandAgilent Technologies
Model5800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals