Maintenance and Troubleshooting
60 Agilent 5800 and 5900 ICP-OES User's Guide
Figure 16B. Flush the gas ports and ball joint connector with deionized water.
4 If the torch was disassembled for cleaning, thoroughly wash the quartz outer
tube-set and the removable top seal. See Figure 17A and Figure 17B.
Figure 17A and B. Rinse the quartz outer tube-set and top seal with deionized water.
5 Dry the torch completely and re-assemble the semi-demountable torch if
necessary, prior to using the torch on the instrument.