Using the CE in Empower
Agilent 7100 CE in Empower Environment 21
Executing Runs
NOTE
The Empower Run Time is shown in the Single Run tab and in the sample sets. The
Agilent Stop Time is an instrument method parameter.
If the Run Time is longer than the instrument method´s Stop Time, the Run Time
is
used for the measurements, regardless of what is stated in the instrument me
thod.
The online plot and the resulting electropherogram end with the Run Time.
If the Run Time is shorter than the instrument method´s Stop Time, the online-plot
stops at the Run Time but the run continues until the end of the method´s Stop
Time. The resulting electropherogram ends at the method´s Stop Time. The longer
time is used for recording data.
The method´s Post Time is independently added after the Stop Time, and no data is
recorded during the post time. The next sequence run starts after the Post Time of the
previous run is finished.