EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7100

Agilent Technologies 7100 Technical Note

Agilent Technologies 7100
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
APPENDIX - Examples
Agilent 7100 CE in Empower Environment 39
Example: Multiple injections using the IQ Kit for CE
For testing multiple injections in one sequence, the instrument method and
the experimental setup described in section Example: Single test run can be
used. See the sequence table (Figure 27) and the overlay of the resulting
electropherograms (Figure 28).
A set of six injections was used to calculate system precision. The peak
areas and relative standard deviations (Excel: STDEV) for the areas,
migration times and the corrected areas are summarized in the following
table.
Table 2 Peak areas and relative standard deviations
Area t
m
[min] t
m
[s] corr. Area (area/t
m
[s])
Injection 1
50049
2.730
163.80
Injection 2
50547
2.723 163.38 3.09383E+02
Injection 3
50189
2.717 163.02 3.07870E+02
Injection 4
49767
2.711 162.66 3.05957E+02
Injection 5
49187
2.706 162.36 3.02950E+02
Injection 6
49579
2.702 162.12 3.05817E+02
Average 49886 2.715 162.89 3.06254E+02
Deviation STDEV: 0.96 % 0.39 % 0.39 % 0.72 %
Figure 27 Sequence table for multiple injections
Sixfold sample injection

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals