Using the CE in Empower
Agilent 7100 CE in Empower Environment 23
Sequence Execution/Sample sets
Empower offers multiple injections from the same vial as well as
consecutive injections from different injection vials. The details have to be
added to the sample/sample set table (sequence), depending on the
required workflow.
1 Click the Sample tab to open the table.
2 Click into the first line to fill in the appropriate values. For the
following values the entries are mandatory:
Vial position
Injection volume
Number of injections
Sample name
Method set
Run time (use the same time as in the instrument method)
NOTE
The parameter Inj. Vol (µl) requires a value. This value is not used for the injection, but it
is reported. The instrument method parameters (for example:
pressure and time) are used
for the injection.
(Instead of volume the quantifiable parameters are pressure × time for hydrodynamic
injection, or voltage × time for electrokinetic injection).
Entering an injection volume of 0 µl does not result in a blank run. The
the run execution, but it is reported.
3 Start the sequence. Reminder: During run execution, the red Stop
button allows you to stop the run at any time.