Removing Covers and Components 2
Maintenance 19
To Remove the Valve/Loop Cover
1 Gather the following:
T- 20 Torx driver
2 Remove the pneumatics front cover. See “To Remove the
Pneumatics Front Cover”.
3 Remove the valve thermal enclosure. See “To Remove the
Valve Thermal Enclosure”.
4 Completely loosen the three T- 20 Torx captive screws on
the valve/loop cover (Figure 11).
Figure 11 Valve/loop cover
5 Remove the valve/loop cover.
Reinstallation is the reverse of these steps.