42 Maintenance
3 Maintenance
To Replace the Sample Loop Adapters
Normally, only change or replace the sample loop adapters
when changing sample loop size. A full set of adapters
shipped with the Headspace.
1 Gather the following:
Lint- free gloves
Sample loop (see Table 7)
Sample loop adapters, as needed (see Table 2)
Needle nose pliers
2 Depending on your hardware configuration, remove the
loop cover or the valve/loop cover, then remove the
sample loop. Follow the steps in “To Replace the Sample
Loop”.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
NOTE
The loop cover is used as an example in this procedure. Steps for the
valve/loop cover are similar.