26 Maintenance
3 Maintenance
To Replace the Sample Tray Seals (12-Vial Model)
The sample tray seals normally do not require replacement.
However, they may need to be cleaned or replaced after
cleaning up spills or other maintenance.
1 Gather the following:
Replacement vial tray seals
2 Lift the rotating top cover.
3 Remove all vials from the carousel.
4 If needed, remove the old vial tray seals.
5 If needed, clean the tray assembly. See “To Clean the
Sample Tray (12- Vial Model)”.
6 Place the new sample tray seals into the vial slots in the
tray.
WARNING
Vials may be hot. Before handling, allow time for vials to cool as
needed. Wear heat resistant gloves as needed.
Rotating top cover