EasyManuals Logo

Agilent Technologies 7820A Operating Guide

Agilent Technologies 7820A
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Operating Basics 2
Operating Guide 27
To Shut Down the GC for More Than a Week
1 Load a GC maintenance method and wait for the GC to
become ready. For more information about creating
maintenance methods, see the Maintaining Your GC
manual. (If a maintenance method is not available, set all
heated zones to 40 °C.)
2 Turn off the main power switch.
3 Shut off all gas valves at the gas source.
4 When the GC is cool, remove the column from the oven
and cap both ends to keep out contaminants.
5 Cap the inlet and detector column fittings and all GC
external fittings.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If they are hot, wear heat-resistant gloves to protect
your hands.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals