Configuration Tasks 6
Operating Guide 61
To restore the default GC IP address
During installation and sometimes during operation, you may
need to reset the GC IP address or change its IP addressing
mode.
To restore the default IP address, press and hold the [Prep
Run] key while power cycling the GC. After reboot, the GC IP
address will return to:
IP address 192.168.0.26
Subnet Mask 255.255.255.0
Gateway 192.168.0.1